EasyManua.ls Logo

YOKOGAWA PH202G (S) - Page 53

YOKOGAWA PH202G (S)
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IM 12B6C3-E-E
Parameter setting 5-16
Code Display Function Function detai• X Y Z Default values
Calibration functions
20 *
t.SEC Stability check time 5 sec.
*PH Stability check pH 0.02 pH
21 *AS.LOW As Pot low limit -120 mV
(As Pot) *AS.HI As Pot high limit 120 mV
21 *ZR.LOW Zero Point low limit 5.00 pH
(Zero) *ZR.HI Zero Point high limit 9.00 pH
22 *SL.LOW Slope low limit 70 %
*SL.HI Slope high limit 110 %
23 *ITP Set ITP Preset calibration data from manufacturer 7.00 pH
(pH) *SLOPE Set slope or from laboratory determinations. 100 %
*ASP.1D Set As Pot For the main parameter 0.0 mV
Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.
*ASP Set As Pot For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.
*ASPmV Set As Pot ORP For parameter 2 (when activated in
service code 02)
23 *ASP.1D Set As Pot (ORP) For the main parameter
(ORP) Press YES to confirm 0.1 mV resolution,
then set value with >, ^, ENT keys.
Press NO to change to *ASP.
*ASP Set As Pot For the main parameter
Press YES to confirm 1 mV resolution,
then set value with >, ^, ENT keys.
24 *BUF.ID Buffer table 4 Buffer tables to NIST (formerly NBS)
25 *BUF.ID Buffer table 7 (see section 10 for table details)
26 *BUF.ID Buffer table 9 User adjustable for special requirements
27 *ZERO.P Enable zero point in Disable zero point (enable As Pot) 0 0 Disabled
pH units Enable zero point (disable As Pot) 1
28-29 Not used
pH
mV
ITP
As pot
0 mV
0 ºC
10 ºC
25 ºC
12B6B3-12
ASYMMETRY POTENTIAL
mV
500
0
- 200
14
pH
ITP
Zero Point
0
7
Fig. 5-1.

Table of Contents

Related product manuals