10-7-7. Changes made by software release 1.7
- Default * T.COEFF changed from 0.00 to -0.00.
- Error 5.1 occurs (instead of 4.1) if no sensor is connected.
- E12 can only occur if second process is ORP or rH.
10-7-8. Changes made by software release 1.8
- Communications with PH201
* B possible.
- Three new temperature sensors (DKK 350, 6K8 and NTC10K).
- Periodically unused errors are reset.
- Service 79 added for loading defaults excepts pH buffer tables.
- No longer PIN needed for communication and “logbook scrolling”.
10-7-9. Changes made by software release 1.9
- Enabling user to set Zero point limits in service code 21.
- mA table handling improved.
- Interpolation mA table improved.
- Communication with PH201
* B improved (WASH).
- High impedance limit raised to 2G (as described in IM).
10-7-10. Changes made by software release 2.0
- E20 is cleared after the programmed data was recovered.
10-7-11 Changes made by software release 3.0
- The maximum ORP span is set to 3000mV (was 2000mV)
Communication is default set to enabled / write enabled
10-7-12 Changes made by software release 3.3
- The
NTC10k
was replaced by the
PTC10k.
10-7-13 Changes made by software release 3.4
- Updated internal tester identification range
.
10-7-14 Changes made by software release 3.5
- Fixed rare HART communication failure
.
10-7-15 Changes made by software release 3.6
- Solve problem with E4.1 / E5.1 impedance errors after loading all parameters from DCS
- Burn low value set to 3.6 mA, only selectable if HART-Communication is disabled
- Prevent incidental reset of the unit while loading default settings
- Prevent incidental communication problems with MH-02 PC-HART modem
10-7-15 Changes made by software release 3.7
- Implementation of Burn low in combination with HART changed.