EasyManuals Logo

YOKOGAWA UT750 User Manual

YOKOGAWA UT750
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
1-12
<Toc> <1. Installation>
IM 05D01B02-41E
Environmental Conditions
Normal operating conditions:
Ambient temperature: 0 to 50C (40C or less for side-by-side close installation)
Temperature change rate: 10C/h or less
Ambient humidity: 20 to 90% RH (no condensation allowed)
Magnetic field: 400 A/m or less
Continuous vibration at 5 to 14 Hz: Full amplitude of 1.2 mm or less
Continuous vibration at 14 to 150 Hz: 4.9 m/s
2
or less
Short-period vibration: 14.7 m/s
2
, 15 seconds or less
Shock: 14.7 m/s
2
or less, 11 ms
Installation height: Height above sea level of 2000 m or less
Warm-up time: 30 minutes or more after power on
Transportation and storage conditions:
Temperature: -25 to 70C
Temperature change rate: 20C/h or less
Humidity: 5 to 95% RH (no condensation allowed)
Effects of changes in operating conditions
- Effects from changes in ambient temperature:
- On voltage or thermocouple input, 1 V/C or 0.01% of F.S./C,
whichever is larger
- On remote input, 0.02% of F.S./C
- On RTD input, 0.05C/C (ambient temperature) or less
- On analog output, 0.05% of F.S./C or less
- Effects from power supply fluctuation (within rated voltage range)
- On remote input, 1 V/10 V or 0.01% of F.S./10 V, whichever is larger
- On analog output, 0.05% of F.S./10 V or less
1st Edition : May 31,2000-00

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA UT750 and is the answer not in the manual?

YOKOGAWA UT750 Specifications

General IconGeneral
BrandYOKOGAWA
ModelUT750
CategoryController
LanguageEnglish

Related product manuals