WARNING! Calibration should always be performed in units of psig.
Calibration of this instrument is for the sensor only (not a leak calibra-
tion) and is factory set. Leak standards can be obtained from Zaxis
Inc. and can be used as a transfer standard to establish applicable leak
rates.
All standards should have at least three tiers of uncertainty. i.e. Isaac
has a tolerance of 0.3% FS (full scale) therefore pressure standards
should be at least equal to or less than 0.1% FS.