25
Zaxis Inc. LLC
801.264.1000
2442 South 2570 West
West Valley City, UT 84119
Calibration Modication
WARNING! If a calibration point data led is pressed unintendedly, the
only way to keep from changing calibration is to immediately shut-off power
to the unit.
To change or ne tune a set point, do the following:
1. Adjust the regulator to the desired calibration point on the pressure standard
2. Press the value eld of the calibration point to be modied.
3. A numeric keypad will appear, enter the value to be set, and select enter.
4. The current reading of the sensor will adjust to the corrections and display the
new value.
5. Exit and save the calibration by pressing the “Back” button.
!
Calibration points should be incremental; point 1 being the
lowest value. Values do not need to be whole numbers.
Special Note
Sequential models can trap a small amount of pressure internally between the channel
selection valve and the pressure transducer. Make sure to open the valves to clear any
pressure before pressing the “TARE” button.