4 ZOLL AED 3 Operator’s Manual
Use the ZOLL AED 3 CPR Uni-padz
TM
defibrillation pads for both
adult and child victims. The ZOLL AED 3 CPR Uni-padz defibrillation
pads may be used on children less than 8 years old or weighing less
than 55 lbs (25 kg) when the Child Selection button has been
pressed.
The ZOLL AED 3 defibrillator is capable of delivering 200 joules. If
necessary, the defibrillator can be completely deactivated by
turning it off and removing the battery pack.
For safety, the semi-automatic version of the ZOLL AED 3
defibrillator automatically disarms a fully charged unit after 30
seconds if the Shock button is not pressed.
Always stand clear of the victim when delivering a shock.
Defibrillation energy delivered to the victim may be conducted
through the victim’s body and cause a lethal shock to those
touching the victim.
DO NOT TOUCH the defibrillation pad surfaces, the victim, or any
conductive material touching the victim during ECG analysis or
defibrillation.
Move the victim away from electrically conductive surfaces prior to
use of equipment.
DO NOT use the ZOLL AED 3 defibrillator near or within puddles of
water.
Keep the victim as motionless as possible during ECG analysis.
DO NOT use the ZOLL AED 3 defibrillator near flammable agents,
such as gasoline, oxygen-rich atmospheres or flammable
anesthetics.
MR Unsafe - Keep the ZOLL AED 3 defibrillator away from magnetic
resonance imaging (MRI) equipment.
Turn off cell phones and two-way radios to avoid radio frequency
interference from high-power sources that might cause the
defibrillator to interpret cardiac rhythms incorrectly.