EasyManua.ls Logo

ZOLL AED 3 - Safety Summary; Warnings

ZOLL AED 3
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ZOLL AED 3 Operator’s Manual 3
Safety Summary
The following section describes general warnings and safety
considerations for administrators, rescuers, and victims.
Warnings
Federal (U.S.A.) law restricts this AED for sale to or on the order of
a physician.
Use the ZOLL AED 3 defibrillator only as described in this manual.
Improper use of the AED can cause death or injury.
DO NOT use or place the defibrillator in service until you have read
the ZOLL AED 3 Operator’s Manual.
DO NOT use or place the ZOLL AED 3 defibrillator in service if the
green check (
) is not displayed in the AED’s status indicator
window (located to the right of the On/Off button). The green check
displays when the AED is ready to use.
DO NOT use or place the ZOLL AED 3 defibrillator in service if the
unit emits a beeping tone when it is off.
DO NOT perform any unauthorized modification to the ZOLL AED 3
defibrillator.
DO NOT disassemble the unit. A shock hazard exists. Refer all
servicing to qualified personnel.
Connect the defibrillation pads cable to the defibrillator BEFORE
installing the battery pack.
Keep the defibrillation pads cable connected to the defibrillator at
all times.
The ZOLL AED 3 defibrillator should only be used by properly
trained individuals.