ZOLL AED 3 Operator’s Manual 23
Self Tests
The ZOLL AED 3 defibrillator performs the following self tests to verify
unit integrity and its readiness for emergency use. See the ZOLL AED 3
Administrator’s Guide for more detailed information about these tests.
Table 2: ZOLL AED 3 Self Tests
Self Test When Performed Functions Performed
Battery Installation Automatically, after battery
pack is installed or
replaced.
Basic functions:
Battery Capacity
Defibrillation Pads
Connection
Defibrillation Pads/
Battery Expiration
Verification
•ECG Circuitry
Defibrillator Charge and
Discharge Circuitry
(2 Joule Test)
Microprocessor
Hardware/Software
CPR Circuitry and Sensor
(if defibrillation pads with
CPR functionality are
connected)
Audio Circuitry
Additional function:
200 Joule Charge Test
Power On Automatically, whenever
the AED is turned on.
Basic functions:
Same as Battery Installation
Self Test basic functions.
Manual By pressing and holding the
AED’s On/Off button for at
least 5 seconds.
Basic functions:
Same as Battery Installation
Self Test basic functions.
Additional function:
200 Joule Charge Test