4.6
TEST
SETUP
«
Remove
0.1Q
resistor
and
replace with
1KQ.
e
Connect
DMM
across
resistor.
e
Set
voltage
scale
to
200
mV.
«
Measure
voltage
across
resistor.
4.7
OFF
LEAKAGE
CURRENT
TEST
«
Measure
across
resistor
with
unit
off.
*
Voltage
should
be
less
than
170
mV
(<170
LA
of
leakage
current.)
Mark
the
Checkout
Procedure
Data
Sheet.
2.