SERVICE
MANUAL
SECTION
7.
DEFIBRILLATOR
TESTS
WARNING
e
Limit
internal
discharges
to
100
Joules
or
less.
Do
not
discharge
the
PD™
1400-series
device
internally
more
than
3
times
in
one
minute.
Multiple
internal
discharges
at
more
than
100
Joules
maydamage
the
unit.
e
All
persons
near
the
equipment
must
be
warned
to
“STAND
CLEAR”
prior
to
defibrillator
discharge.
Perform
this
Action:
Look
for:
Circle
the
Result
7.1
Disconnect
external
"PADDLE
FAULT"
message
is
|
YES
NO
paddles.
Set
SELECTOR
|
displayed
on
CRT.
SWITCH
to
DEFIB
ON.
7.2
Connect
external
defib-
"PADDLE
FAULT"
message
YES
NO
rillator
paddles
to
the
unit.
|
replaced
by
"PADDLES".
Set
energy
level
to
100].
Charge
time
greater
than
2
YES
NO
Make
sure
paddles
are
seconds.
firmly
in
storage
wells.
Press
CHARGE
button.
7.3
Press
Apex
paddle
"100)
ВРУ"
message
YES
NO
CHARGE
bution.
displayed.
Press
only
Apex
"1007
RDY"
message
YES
NO
DISCHARGE
button
displayed;
unit
does
not
discharge.
YES
NO
Press
only
Sternum
"100]
RDY"
message
displayed
DISCHARGE
button
unit
does
not
discharge.
7.4
Press
both
DISCHARGE
|
Unit
discharges.
YES
NO
buttons
"100j
READY"
message
YES
NO
disappears.
"TEST
OK"
message
appears.
|
YES
NO
"TEST
OK"
prints
on
strip
YES
NO
chart.