PROCEDURES
SECTION
7.
DEFIBRILLATOR
TESTS
(CONTINUED)
Perform
this
Action:
Look
for:
Circle
the
Result
Set
energy
level
to
150]
"1507
RDY"
message
YES
NO
Press
CHARGE
button
displayed.
and
time
charge.
Press
DISCHARGE
"150J
RDY"
message
YES
NO
buttons.
disappears.
Unit
discharges.
Baseline
return
in
<=
4
sec.
YES
NO
Energy
delivered
to
defib
YES
NO
tester
127.5
-
172.5).
Delay
time
displayed
on
Defib
|
YES
NO
analyzer
<
50
ms.
YES
NO
Set
energy
level
to
2005
"200
RDY"
message
Press
CHARGE
button
displayed.
and
time
charge.
YES
NO
Press
DISCHARGE
“200]
RDY"
message
buttons.
disappears.
Unit
discharges.
YES
NO
Baseline
return
in
<=
4
sec.
YES
NO
Energy
delivered
to
defib
tester
180
-
2307.
:
:
|
YES
NO
Delay
time
displayed
on
Defib
analyzer
<
50
ms.
Set
energy
level
to
300]
o
Press
CHARGE
button
"300J
RDY"
message
YES
N
and
time
charge.
displayed.
Press
DISCHARGE
buttons.
"3007
RDY"
message
YES
NO
disappears.
Unit
discharges.
Baseline
return
in
<=
4
sec.
YES
NO
Energy
delivered
to
defib
YES
NO
tester
255
-
345].
Delay
time
displayed
on
Defib
YES
NO
analyzer
<
50
ms.
1-37