SERVICE
MANUAL
PD™
1400-SERIES
TEXT
MESSAGES
Text
Messages
Description
PACE/DEFIB
FAULT
Pace/Defib
Board
is
Missing
or
Board
Revision
is
Unsupported.
SET
CLOCK
Real-Time
Clock
(RTC)
Failure
(Bad
Date/time)
or
RTC
back-up
power
supply
Failure.
REPORT
FAULT
EEPROM
write
failure
or
EEPROM
Erase
Failure.
REPORT
FULL
Cannot
store
more
Summary
Report
-
Data
because
the
EEPROM
is
full.
ERASING
REPORT
Summary
Report
storage
being
erased.
USER
SETUP REQ
Both
copies
of
the
calibration
or
configuration
data
are
bad
and
default
values
must
be
used.
LOW
BATTERY
Battery
pack
voltage
is
below
the
|
absolute
minimum
or
an
acceptable
threshold.
PACER
LEAD
OFF
Pace
leads
are
off.
PADDLE
FAULT
Paddles
are
selected
but no
Paddles/MFE
are
connected
or
they
are
only
partially
connected.
NO
PAPER
No
Strip
Chart
Recorder
Paper
ECG
LEAD
OFF
ECG
Leads
Off.
BUTTON
ERROR
Too
many
buttons
pressed
simultaneously.
SHORTED
DISCH
Dumped
more
than
200]
into
internal
DEFIB
self-test
shorting
bar.
OPEN
ATR
DISCH
Capacitor
voltage
still
high
when
relay
returns.
(Open
air
discharge)