SERVICE
MANUAL
PD™
1400-SERIES
TEXT
MESSAGES
(CONTINUED)
Text
Messages
Description
DISABLE
SYNC
(PD™
2000/D
2000
only)
Displayed
if
the
ANALYZE
button
is
pressed
and
the
device
isin
SYNC
DEFIB
|
mode.
The
unit
should
be
taken
out
of
SYNC
mode
by
pressing
the
SYNC
button
before
attempting
to
activate
Advisory
Analysis
ANALYSIS
HALTED
(PD™
2000/D
2000
only}
Analysis
was
halted
due
to
either
an
ECG
that
was
detected
as
too
noisy
or
due
to
the
operational
mode
of
the
unit
being
switched
during
analysis
VF
ALARMS
OFF
(PD™
2000/D
2000
only)
This
indicates
that
the
advisory
analysis
will
not
be
performedwhen
Heart
rate
alarms
are
enabled
while
in
pace
mode
SEL.
ELECTRODES
(PD™
2000/D
2000
only)
Displayed
if
the
ANALYZE
button
is
pressed
and
the
device
is
in
SYNC
DEFIB
mode.
The
unit
should
be
taken
out
of
SYNC
mode
by
pressing
the
SYNC
button
ECG
TOO
LARGE
ADJUST
ECG
GAIN
If
the
ECG
signal
is
too
large
during
advisory
analysis,
this
message
will
prompt
the
user
to
decrease
the
ECG
gain.
NOISY
ECG
RETRY
ANALYSIS
This
message
indicates
that
the
advisory
analysis
algorithm
determined
that
the
ECG
signal
was
too
noisy
to
properly
determine
the
rhythm
characteristics.
DI-4