NOTICE
To reduce risk of damage and ensure an accurate result
with the 3M™CleanTrace™ LuminometerLX25:
Do not operate the Clean-Trace LuminometerLX25in
close proximity to strong magneticfields.
Do not store or operate the Clean-Trace
LuminometerLX25at temperatures above70°C.
Do not store or operate the Clean-Trace
LuminometerLX25at temperatures below -40°C as
condensation may occur. The instrument should be
allowed to warm to room temperature before use
and a cloth or tissue should be used to wipe dry the
outersurfaces.
Only insert 3M™Clean-Trace™ ATP Tests into the
Clean-Trace LuminometerLX25chamber.
Do not use cleaning agents in the Clean-Trace
LuminometerLX25chamber.
Ensure the ATP Test is inserted into the Test sleeve and
activated before inserting the Test into the Clean-Trace
LuminometerLX25chamber.
Do not tilt the Clean-Trace LuminometerLX25while
ATP Tests are in thechamber.
Do not leave a used (activated) ATP Test in the
Clean-Trace LuminometerLX25chamber after the
result has been read andrecorded.
Ensure annual recommended service and calibration of
the Clean-Trace LuminometerLX25is completed by the
3M Service Center or 3M authorized servicepersonnel.
Periodically inspect the Clean-Trace
LuminometerLX25for damage, including cracks
or swelling.
Do not store or use the Clean-Trace
LuminometerLX25or power supply in a location
exposed to steam, high humidity, dust, or
strong vibration.
Reference the Environmental Operating Conditions
table in the Important Safety Information for
Clean-Trace LuminometerLX25in this UserManual.
Use only 3M Clean-Trace ATP Tests and 3M accessories
supplied by your local 3M sales representative
ordistributor.
Do not use the Clean-Trace LuminometerLX25if liquid
or a foreign object gets inside thedevice.
Environmental Operating Conditions
Environmental Condition Operating Condition Units
Indoor Use only - Only use the data connections to a
computer that conform to IEC/EN/UL/
CSA60950-1
Notapplicable
- Keep Clean-Trace Luminometer LX25
away from liquids and high humidity.
Altitude 3000maximum Meters
Operating Temperature Range +5to +40(+41to +104) °C(°F)
Relative Humidity 10– 90Non-condensing %
Storage Temperature Range -40to +70(-40to +158) °C(°F)
AC voltage for powersupply 100-240 Volts
Mains supply voltage fluctuations up to +10% of
nominal voltage
Temporary overvoltages occurring on the mains supply
Frequency 50-60 Hertz
AC Current for power supply 300 mA
Power supply output voltage 5 VoltsDC
Power supply output current 1.2 Amps
Input current 850 mAMPs
Overvoltage Category II Notapplicable
Pollution Degree 2 Notapplicable

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3M Clean-Trace Luminometer LX25 and is the answer not in the manual?

3M Clean-Trace Luminometer LX25 Specifications

General IconGeneral
Brand3M
ModelClean-Trace Luminometer LX25
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals