34
Entering the density of a liquid (ldin = 0)
1. Press the UNITS key to select
SG
.
2. Press and hold the UNITS key to display the liquid
density input mode. The display will show the water
temperature currently set (factory setting: 25 °C).
3. Using the RE-ZERO (to increment the value) and
SAMPLE keys (to shift the selected digit), set the
value and press the PRINT key to store.
To cancel the setting procedure and return to the
density measuring mode, press the UNITS key.
The input value is not stored.
The relation between the water temperature and
density is shown below.
°C
+0 +1 +2 +3 +4 +5 +6 +7 +8 +9
0
0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10
0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20
0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30
0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40
0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
50
0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371
60
0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834
70
0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242
80
0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600
90
0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906
Entering the density of a liquid directly (ldin = 1)
1. Press the UNITS key to select
SG
.
2. Press and hold the UNITS key to display the liquid
density input mode. The display will show the liquid
density currently set (factory setting: 1.0000 g/cm
3
)
3. Using the RE-ZERO (to increment the value) and
SAMPLE keys (to shift the selected digit), set the
value and press the PRINT key to store.
To cancel the setting procedure and return to the
density measuring mode, press the UNITS key. The
input value is not stored.
d 1.0000
SG
t---25
SG
SG indicato
r
0.000
SG