27
8.5.
Function List
Class Item
Param-
eter
Description
func
poff
Auto power-off
0
uto powe
-off disabled
Automatically power OFF
1
uto powe
-off enabled
rng
Weighing range
0 Manual range Function to change the
weighing range
(EJ-54D2 only)
1 Automatic range
Cond
Response
0 Fast / sensitive
Software filtering 1
2 Slow / stable
5t-b
Stability band width
0 Judge precisely (±0.5d)
The STABLE indicator
turns on when the
fluctuation interval of the
weighing display is less
than the reference value.
1
(±1d)
2 Judge moderately (±2d)
trc
Zero tracking
0 Disabled
Function to keep the
display value at zero by
tracking the zero point
1 Enabled
pnt
Decimal point
0 Point (.)
Decimal separator
1 Comma (,)
prt
Data output mode
0 Command and stream modes
Auto-print A: + data
Auto-print B: +/- data
1 Command mode and PRINT ke
2
Command mode, PRINT key
and auto-print A
3
Command mode, PRINT key
and auto-print B
4 Command mode onl
pU5e
Data output pause
0 No pause (general equipment)
Interval between
continuous data
1 1.6 seconds (for AD-8121B)
info
GLP output
0 No output
GLP output format 1
D-8127/AD-8121B format
*
2 General format
bp5
Baud rate
0 2400 bps
1 4800 bps
2 9600 bps
3 1200 bps
btpr
Data and parity
0 7 bits, even parity
1 7 bits, odd parity
2 8 bits, non parity
aCai
ACAI function
0 ACAI disabled
When “0” is set, no additional
samples required.
1 ACAI enabled
Umin
Minimum unit weight
0 1d
d = readability (scale
interval or “division”)
1 1/10d
2 Total sample weight 5d
(
**
)
5mpl
Number of samples
0 10 pcs
The number of samples shown first when
entered into the unit weight storing mode
1 25 pcs
2 50 pcs
3 100 pcs
4 5 pcs
: Factory setting