15
Return to
top of page
% Slope Value
pH CALIBRATION
–––––
pH Check Value
Continued from previous page.
l000
.
700
.
6 OPERATION…
…6.2.1 Operating Page
Slope Value:
a value between the programmed minimum % slope value (see
pH Calibration Minimum
Slope Value
in the Set Up Parameters Page on page 25) and 105% should be displayed.
If the value is outside these limits the electrode system must be checked.
pH Check Value:
is displayed as an additional indication of electrode system condition; 7pH being the
optimum value for glass electrodes and 0pH for Antimony electrodes.
Advance to PH Calibration Page – see Sections 6.2.2 to 6.2.4.