EasyManua.ls Logo

ABB 4630 - Page 34

ABB 4630
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
Return to
top of page
.
Monitoring pH
700
Redox (ORP)
250
or
Alter Fact. Code
00000
Adjust RTX Zero2
–––––
Adjust RTX Span2
–––––
Continued from previous page.
Adjust RTX Zero
–––––
Adjust RTX Span
–––––
8 CALIBRATION
8.3 Factory Settings Page
Adjust Retransmission Zero
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the Set Up Retransmission Page does
not affect the reading.
Advance to next parameter.
Adjust Retransmission Span
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the Set Up Retransmission Page does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Zero (available only on 4630/800, 4631/800, 4635/800 and
4636/800 instruments)
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the Set Up Retransmission Page does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Span (available only on 4630/800, and 4631/800, 4635/800
and 4636/800 instruments)
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the Set Up Retransmission Page does
not affect the reading.
Alter Factory Settings Access Code
Set the factory settings access code to a value between 00000 and 19999.
Return to Operating Page.

Related product manuals