Installing the sensor head 17
5 Once fully tightened, make sure that the sensor is in horizontal position with the sampling access
panel pointing upwards. Rotate the sensor as required to achieve this.
NOTICE
Failure to position the CoreSense horizontally, with the access panel pointing upwards,
will cause the thermal pump to malfunction and may result in erroneous readings.
6 With the 5 mm Allen key, tighten the rotating flange nuts to a torque of 8.1 N-m (6.0 lb-ft) according
to the following tightening pattern.
5
6
3
4
2
7
1
8
Start running in all the bolts or nuts a few threads and then run them hand-tight. Tighten by hand,
then according to the torque specification.
NOTICE
Do not over tighten.