EL3000 CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. C 19
Test gases – Limas23
calibration
-point calibration
with calibration cells
2
or air or UV sample component-free gas
Calibration cells or test gas for each sample
component
without calibration cells
2
or air or UV sample component-free gas
Test gas for each sample component
without calibration cells
2
or air or UV sample component-free gas
Test gas for each sample component
/ Magnos28 or oxygen
with calibration cells
(automatic calibration, i.e. Magnos206
/ Magnos28
-point calibration)
2
or oxygen and UV sample component-free gas
Either calibration cells and test gas for oxygen
detector
or test gas for each sample component or
/ Magnos28 or oxygen
without calibration cells
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
/ Magnos28 or oxygen
without calibration cells
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
Dew point
The dew point of the test gases must be approximately equal to the dew point of the sample gas.
Note
The instructions for calibration must be observed, see Operating instruction.