EL3000 CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. C 19
Test gases Limas23
Analyzer(s)
Test gas for the zero
calibration
Test gas for the end
-point calibration
Limas23
with calibration cells
(automatic calibration)
N
2
or air or UV sample component-free gas
Calibration cells or test gas for each sample
component
Limas23
without calibration cells
(automatic calibration)
N
2
or air or UV sample component-free gas
Test gas for each sample component
Limas23
without calibration cells
(manual calibration)
N
2
or air or UV sample component-free gas
Test gas for each sample component
Limas23 + Magnos206
/ Magnos28 or oxygen
sensor
with calibration cells
(automatic calibration, i.e. Magnos206
/ Magnos28
with single
-point calibration)
N
2
or oxygen and UV sample component-free gas
Either calibration cells and test gas for oxygen
detector
or test gas for each sample component or
for each detector
Limas23 + Magnos206
/ Magnos28 or oxygen
sensor
without calibration cells
(automatic calibration)
N
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
Limas23 + Magnos206
/ Magnos28 or oxygen
sensor
without calibration cells
(manual calibration)
N
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
Dew point
The dew point of the test gases must be approximately equal to the dew point of the sample gas.
Note
The instructions for calibration must be observed, see Operating instruction.