24 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Test gases for the calibration
Uras26
Analyzer Test gas for zero-point calibration Test gas for end-point calibration
Uras26 with calibration cells
(automatic calibration)
N
2
or air or IR sample component-free
gas
– (calibration cells)
Uras26 without calibration
cells (automatic calibration)
N
2
or air Span gas*
Uras26 without calibration
cells (manual calibration)
N
2
or air Test gas for each sample component
or for each detector
Uras26 + Magnos206
(automatic calibration,
i.e. Magnos206 with
single-point calibration)
IR sample component-free test gas
with O
2
concentration in an existing
measuring range or ambient air. Same
moisture content as process gas.
Calibration cells or span gas*
Uras26 + Magnos206
(manual calibration)
Zero reference gas for Uras26 or
Magnos206, or IR sample compo-
nent-free test gas with O
2
concentra-
tion in an existing measuring range
or ambient air. Same moisture con-
tent as process gas.
Span gas for all sample components
in the Uras26 and Magnos206 (possi-
bly only for the Uras26 if a sin-
gle-point calibration is carried out for
the Magnos206)
Uras26 + Magnos28
(automatic calibration,
i.e. Magnos28 with
single-point calibration)
IR sample component-free test gas
with O
2
concentration in an existing
measuring range or ambient air. Same
moisture content as process gas.
Calibration cells or span gas*
Uras26 + Magnos28
(manual calibration)
Zero reference gas for Uras26 or
Magnos28, or IR sample compo-
nent-free test gas with O
2
concentra-
tion in an existing measuring range
or ambient air. Same moisture con-
tent as process gas.
Span gas for all sample components
in the Uras26 and Magnos28 (possi-
bly only for the Uras26 if a sin-
gle-point calibration is carried out for
the Magnos28)
Uras26 + Caldos27
(automatic calibration,
i.e. Caldos27 with
single-point calibration)
IR sample component-free test gas
with a known and constant rTC value
(possibly also dried room air)
Calibration cells or span gas*
Uras26 + Caldos27
(manual calibration)
Zero reference gas for Uras26 or
Caldos27, or IR sample component-
free test gas with a known rTC value
Span gas for all sample components
in the Uras26 and Caldos27 (possibly
only for the Uras26 if a single-point
calibration is carried out for the Cal-
dos27)
Uras26 + Caldos25
(automatic calibration)
Sample component-free test gas or
substitute gas for Uras26 and
Caldos25
Test gas or substitute gas mixture
for all sample components in the
Uras26 and in the Caldos25*
Uras26 + Caldos25
(manual calibration)
IR sample component-free test gas
for Uras26 and sample component-
free test gas or substitute gas for
Caldos25
Span gas for all sample components
in the Uras26 and test gas or substi-
tute gas with known sample compo-
nent concentration for Caldos25
* Test gas mixture for multiple sample components possible if no or negligi-
ble cross-sensitivity is present

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals