EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 25
Magnos206
Analyzer Test gas for zero-point calibration Test gas for end-point calibration
Magnos206 Oxygen-free process gas Process gas with a known O
2
concen-
tration
Magnos206 with a sup-
pressed measuring range
Test gas with O
2
concentration near
the starting point of the measuring
range
Test gas with O
2
concentration near
the end point of the measuring range
Magnos206 with single-
point calibration
Test gas with O
2
concentration in an
existing measuring range or ambient
air. Same moisture content as pro-
cess gas.
Magnos206 with substitute
gas calibration
Oxygen-free process gas or substi-
tute gas (O
2
in N
2
)
Substitute gas, e.g. dried air
Magnos28
Analyzer Test gas for zero-point calibration Test gas for end-point calibration
Magnos28 Oxygen-free process gas Process gas with a known O
2
concen-
tration
Magnos28 with a sup-
pressed measuring range
Test gas with O
2
concentration near
the starting point of the measuring
range
Test gas with O
2
concentration near
the end point of the measuring range
Magnos28 with single-point
calibration
Test gas with O
2
concentration in an
existing measuring range or ambient
air. Same moisture content as pro-
cess gas.
Magnos28 with substitute
gas calibration
Oxygen-free process gas or substi-
tute gas (O
2
in N
2
)
Substitute gas, e.g. dried air
Caldos27
Analyzer Test gas for zero-point calibration Test gas for end-point calibration
Caldos27 Sample component-free test gas or
process gas
Test gas or process gas with a known
sample component concentration
Caldos27 with a suppressed
measuring range
Test gas with a sample component
concentration near the starting point
of the measuring range
Test gas with a sample component
concentration near the end point of
the measuring range
Caldos27 with single-point
calibration
Test gas with a known and constant
rTC value (standard gas; possibly also
dried room air)
Caldos25
Analyzer Test gas for zero-point calibration Test gas for end-point calibration
Caldos25 Sample component-free test gas or
process gas
Test gas or process gas with a known
sample component concentration
near the end point of the measuring
range
Caldos25 with substitute
gas calibration
Sample component-free substitute
gas
Substitute gas with a known sample
component concentration near the
end point of the measuring range

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals