EasyManuals Logo

ABB EasyLine EL3060 Series User Manual

ABB EasyLine EL3060 Series
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 55
Set the IP address in the gas analyzer
Setup Device settings Ethernet
Parameters
The parameters which have to be input depend on the DHCP setting:
DHCP on: Network name (max. 20 characters, no blanks or special characters),
DHCP off: IP address, IP address mask and IP gateway address.
The network name can only be altered in the configurator (see page 53). The
default network name consists of "EL3K" and the last six positions of the
MAC address (example: "EL3KFF579A").
If the parameter "DHCP" is set to "off", the Ethernet configuration is reset to
the standard configuration (default IP address); this will prevent the inad-
vertent assignment of an IP address from a DHCP pool.
NOTE
The parameter "DHCP" must be set to "off" when the gas analyzer is not
connected to a network. This prevents the gas analyzer from constantly
trying to connect to a network.
Addresses
The IP address, IP address mask and IP gateway address must be obtained
from the system administrator.
NOTE
The address bits that can be varied in the address mask may not all be set to
0 or 1 (broadcast addresses).
MAC address
The 12-digit MAC address is unique and stored in each device during manu-
facture. It cannot be altered.

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals