68 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Manual calibration: Execution
Menu path
Operation Calibration Manual Calibration
Wait until the warm-up phase has ended
The gas analyzer may not be calibrated until the warm-up phase has been
concluded.
Duration of the warm-up phase:
Uras26: approx. ½ hour without, approx. 2.5 hours with thermostat
Magnos206: approx. 2 hours
Magnos28: approx. 2 to 4 hours
Caldos27: approx. ½ hour
Caldos25: 1 to 4 hours, depending on the measuring range
NOTES
A zero-point calibration must always be carried out before an end-point
calibration. A zero-point calibration can also be carried out on its own.
Manual calibration cannot be carried out when an automatic calibration is
running.