Endura AZ20 Series Probe 5 Dismantling and Reassembly – Probe
IM/AZ20M–EN Rev. B 35
5.12 Replacing the PT1000 Temperature Sensor
Before fitting the PT1000 sensor:
Ensure a PT1000 sensor is available – refer to Section 7.8,
page 44.
Remove the probe from its mounting as detailed in
Section 5.4, page 14 (integral probe) or Section 5.5, page
14 (remote probe).
Remove the probe end cap as detailed in Section 5.6.1,
page 15.
Referring to Fig. 5.20:
1. Loosen the 2 ACJC connection terminal block screws
A
(violet and grey terminal label) and remove existing
PT1000 sensor
B.
2. Connect the new PT1000 to the terminal block and
tighten the 2 ACJC terminal block screws A.
3. Refit the probe end cap – refer to Section 5.6.2, page 15.
4. Re-install the probe as detailed in the probe User Guide
(IM/AZ20P–EN).
Fig. 5.20 Removing / Replacing the PT1000 Sensor