Combustion Oxygen Monitor
Endura AZ20 Series Probe 5 Dismantling and Reassembly – Probe
34 IM/AZ20M–EN Rev. B
5.11 Replacing / Fitting Restrictors
Before proceeding:
Ensure a restrictor kit is available – refer to Section 7.8,
page 44.
Remove the probe from its mounting as detailed in
Section 5.4, page 14 (integral probe) or Section 5.5, page
14 (remote probe).
Referring to Fig. 5.19:
1. Remove the reference air (Ref. Air) fitting A and external
test gas (TG1, TG2) fittings B from the probe body inlets
C using a 14 mm A/F spanner (not supplied).
2. If fitted, remove existing filter pads
D from the probe
body TG1, TG2 and Ref. Air inlets.
3. If fitted, remove existing test gas restrictors E from the
probe body TG1 and TG2 inlets using the 3 mm A/F
hexagon key supplied in the toolkit with the probe.
4. If fitted, remove the reference air restrictor
F from the
probe body (Ref. Air) inlet using the 2.5 mm A/F hexagon
key supplied with the restrictor kit.
5. Remove any O-rings from the new restrictors and discard
(the seal between the restrictors and probe body inlets is
made by the metal-to-metal faces of the restrictors /
inlets).
6. Fit the 2 new (larger) test gas restrictors
G into the TG1
and TG2 inlets on the casting and tighten fully using the
3 mm A/F hexagon key.
Do not use thread sealant (the metal to metal seal is
sufficient).
7. Fit the new (smaller) reference air restrictor H into the
Ref. Air inlet and tighten fully using the 2.5 mm A/F wrench
supplied.
Do not use thread sealant (the metal to metal seal is
sufficient).
8. Fit new filter pads
I in front of each restrictor.
9. Refit the reference air (Ref. Air) fitting
A and external test
gas (TG1, TG2) fittings
B fitting using a 14 mm A/F
spanner.
Ensure fittings are leak-tight, use PTFE tape on the
threads if necessary.
10. Mark the Commissioning label
J on the probe to show
that restrictors are fitted.
11. Re-install the probe as detailed in the probe User Guide
(IM/AZ20P–EN).
12. Make gas supplies to the probe and electrical supplies to
the transmitter – refer to the transmitter User Guide
(IM/AZ20E–EN) for electrical supplies and the probe User
Guide (IM/AZ20P–EN) for gas and reference air supplies.
Note. The restrictors have extremely small orifices so
ensure this procedure is performed in a clean area.
Fig. 5.19 Replacing / Fitting Restrictors
B
A
C
B
I
G
H
F
E
TG2
Ven t
D
J