EasyManua.ls Logo

ABB Endura AZ20 series

ABB Endura AZ20 series
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Combustion Oxygen Monitor
Endura AZ20 Series Probe 3 Troubleshooting – Quick Checks
IM/AZ20M–EN Rev. B 7
3.3 Checking the Heater Resistance
Referring to Fig. 3.3:
1. At the 9-way terminal block A, measure the resistance
across the heater leads B (brown and blue).
2. The measured resistance should be within the ranges
shown in Table 3.1:
3. If the measured resistance is out of range, check the
heater wiring and, if necessary, replace the heater
assembly – refer to Section 5.9, page 24.
If the measured resistance is within range, proceed to
Section 3.4.
3.4 Checking the Thermocouple / Electrode
Assembly Resistance
Referring to Fig. 3.4:
1. Ensure the probe is at ambient temperature (–20 to 70 °C
[–4 to 158 F]).
2. At the 9-way terminal block
A, check the thermocouple /
electrode assembly terminals
B (TC+ green and TC–
white) for open- or short-circuit.
3. Check that the thermocouple / electrode assembly is
isolated from the probe body earth. If it is not isolated,
check for short-circuits to earth in the probe’s internal
wiring.
4. The resistance measured should be within the following
range:
5. If the measured resistance is out of range, replace the
thermocouple / electrode assembly – refer to Section 5.8,
page 20.
If the measured resistance is within range, proceed to
Section 3.5.
Fig. 3.3 Heater Connections
Probe Type Resistance
AZ20 Standard Probes
0.5 to 4.0 m (1.64 to 13.12 ft)
185 to 220
AZ20/ZFG2 Version 25 to 37
Table 3.1 Heater Resistance by Probe Type
B
A
Fig. 3.4 Thermocouple / Electrode Assembly
Connections
Probe Length Resistance
0.5 to 2.0 m (1.64 to 6.6 ft) 10 to 17
2.5 to 4.0 m (8.2 to 13.1 ft) 17 to 23
Table 3.2 Thermocouple / Electrode Assembly
Resistance by Probe Length
B
A

Table of Contents

Other manuals for ABB Endura AZ20 series

Related product manuals