EasyManua.ls Logo

ABB IRB 640 - Page 89

ABB IRB 640
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Repairs
46 Product Manual IRB 640
6. Align the pin and tool with a sliding calliper. See Figure 9.
Figure 9 Aligning the pin and tool with a sliding calliper for axis 1.
Before calibrating axes 2, 3 and 6, calibrate the sensors against each other, using a reference
plane surface, in the same direction. See Figure 10.
Figure 10 Calibrating the sensors.
Axes 2, 3 and 6
7. Release the enabling device.
8. Mount sensor fixture 6808 0011-GM on the bases reference plane, see Figure 11.
9. Mount elbow fixture 6808 0011-LP on the lower arms calibration plane, see Figure 11.
10. Mount sensor fixture 6808 0011-GM at the back of the upper arm, see Figure 11.
11. Mount sync fixture 3HAC 1904-1 on the frame for axis 6 and turning disc,
see Figure 12.
12. Mount inclination instrument 6807 081-D. One sensor should be mounted on the refer-
ence plane and the other on the elbow fixture for axis 2. Both sensors should be posi-
tioned in the same direction. See also Figure 11.
Note that the sensor unit must always be mounted on top of the fixture.
MANIPULATOR
6896 0011-YM
Sliding calliper
0000
Reference plane
Sensor

Table of Contents

Related product manuals