LST300 | COMPACT ULTRASONIC LEVEL TRANSMITTER | LST300EN REV. E
51
5 Select Wet Calibrate Upper to perform a sensor
measurement of the upper calibration point, and then click
OK
6 Confirm the measurement in the Upper Sensor Point set.
7 Ensure the LST300 is measuring the lower calibration point,
by ensuring the conditions as indicated in figure 27 “Lower
calibration point measured”.
Figure 31 Lower calibration point measured
Lower Level Point: The actual short range level
8 Select Wet Calibrate Lower to perform a sensor
measurement of the lower calibration point, and then click
OK.
9 Confirm the measurement in the Lower Sensor Point set.
The level wet calibration is immediately active on the
LST300.