44 LST300 |  COMPACT ULTRASONIC LEVEL TRANSMITTER  |  LST300–EN REV. E
Configure the volume unit
1  Select Calibrate from the main menu.
2  Select Units Config from the “Diagnostics” submenu.
3  Select Volume Unit from the “Unit Config” submenu.
4  Configure the Volume Unit as required (liter is used in this 
example) and then press Next.
Configure the linearization table
1  Select Calibrate from the main menu.
2  Select Linearization from the “Calibrate” submenu.
3  Set Max Volume from the “Linearization” submenu.
In this example, 1832 liters is set as the maximum volume (unit 
already set in section “Configuring the volume unit”).
4  Enter the Linearization Table from the “Linearization” 
submenu.