46 LST300 | COMPACT ULTRASONIC LEVEL TRANSMITTER | LST300EN REV. E
Open channel flow on DTM
1 Click Configuration > Device Setup > Basic Setting.
2 Select
Flow
in the Operate Mode field and click Set.
3 Click Configuration > Open Channel Flow > Open Channel
Selection.
4 Select a value from the Channel Shape box and click Apply.
The
Triangular Notch Thin Plate Weir
is selected here as an
example.
5 The parameters and an image for the selected channel shape are displayed.
6 Complete the parameters and click Calculate.
7 The maximum flowrate is calculated and displayed at the top of the interface.
Note
The max flowrate calculation must be executed before the linearization table generation, to prevent an incorrect linearization
table.