LST300 |  COMPACT ULTRASONIC LEVEL TRANSMITTER  |  LST300–EN REV. E
45
5  Press   to select the position to be changed to the 
linearization point number. The first point to update in this 
example is point 
00
.
6  Press   to change the position to Lev.
7  Use   to enter the editing menu.
8  Set the level for point 00 in the editing menu.
In this example level 
0.000
 m is set.
9  Use   to select OK and return to the “Linearization Table” 
menu.
10 Press 
11 Use   to enter the editing menu.
12 Set the level for point 00 in the editing menu.
In this example a percentage of 
 is set for point 00.
13 Use 
 to select OK and return to the “Linearization Table” 
menu.
14 Repeat steps 5 to 13 for linearization point 01 and 02 
according to the data in the table below:
Point Level %
  
  
  
Open channel flow
The LST300 open channel flow function calculates the flowrate 
in an open channel by measuring level and converting it to a 
flowrate using standard equations. The automated calculation 
of parameters can only be done using DTM or EDD. Alternatively 
the calculation can be done offline, and the setup can be done 
in the field on the HMI using the pre-calculated maximum flow 
value and linearization table.
The open channel flow is not calculated directly. A linearization 
table and the maximum flowrate are calculated and 
downloaded to the device using DTM or EDD. Then this 
linearization table is used on the device to calculate the 
flowrate.
LST300 supports the following 12 channels:
•  Rectangular thin plate weir
•  Triangular notch thin plate weir
•  Rectangular broad crested weir
•  Rectangular throated flume
•  Trapezoidal throated flume
•  U throated flume
•  Round nose horizontal broad crested weir
•  Parshall flume
•  Palmer bowlus flume
•  Round pipe
•  Leopold lagco flume
•  Cut throat flume
Any custom or special channels with available linearization 
transfer curve can also be configured manually.