EasyManua.ls Logo

ABB ProcessMaster 300 - Page 34

ABB ProcessMaster 300
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ex relevant specifications for operation in Div. 2
32 - EN FEX300, FEX500 SM/FEX300/FEX500/FM/CSA-EN
Table 2: Fluid temperature (Ex data) for ProcessMaster models FEP315, FEP515 and HygienicMaster models FEH315,
FEH515
Nominal diameter
Design
Temperature
class
Ambient temperature
- 20 °C ... + 40 °C - 20 °C ... + 50 °C - 20 °C ... + 60 °C
- 40 °C ... + 40 °C
1)
- 40 °C ... + 50 °C
1)
- 40 °C ... + 60 °C
1)
Not thermally
insulated
Thermally
insulated
Not thermally
insulated
Thermally
insulated
Not thermally
insulated
Thermally
insulated
Gas
Gas &
dust
Gas
Gas &
dust
Gas
Gas &
dust
Gas
Gas &
dust
Gas
Gas &
dust
Gas
Gas &
dust
ProcessMaster DN 3 ... DN 2000
HygienicMaster DN 3 ... DN 100
NT
T1
130 °C 130 °C - - - - - - 130 °C
100 °C
2)
110 °C
3)
- - - - - - 80 °C 40 °C - - - - - -
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
NT
T2
130 °C 130 °C - - - - - - 130 °C
100 °C
2)
110 °C
3)
- - - - - - 80 °C 40 °C - - - - - -
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
NT
T3
130 °C 130 °C - - - - - - 130 °C
100 °C
2)
110 °C
3)
- - - - - - 80 °C 40 °C - - - - - -
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
NT
T4
130 °C 130 °C - - - - - - 130 °C
100 °C
2)
110 °C
3)
- - - - - - 80 °C 40 °C - - - - - -
HT 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 40 °C 130 °C 40 °C
1) Low-temperature version (option)
2) Temperature values for ProcessMaster
3) Temperature values for HygienicMaster
NT: standard sensor design, T
medium
maximum 130 °C (266 °F)
HT: high-temperature sensor design, T
medium
maximum 180 °C (356 °F)
Not thermally insulated: The flowmeter sensor is not surrounded by pipe insulation material.
Thermally insulated: The flowmeter sensor is surrounded by pipe insulation material.
Note
The standard version includes explosion protection for gases and dust. Explosion protection for dust is only available for devices featuring a
transmitter in a dual-compartment housing.
If the installation location for the device is classified as a potentially explosive area for gases and dust, the temperature data
in the "Gas & dust" columns in the table must be taken into consideration.
If the installation location for the device is classified as a potentially explosive area for gases only, the temperature data in
the "Gas" column in the table must be taken into consideration.