EasyManua.ls Logo

ABB XFC G5 - Analog Input Calibration (Voltage Mode)

ABB XFC G5
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
132 | XSE RIES G5 | 21060 2 6MN AA
Figure 11-4: XRC
G5
AIs
Figure 11-5: XRC
G5
AIs and mode selector jumpers (default setting)
To set the AI to voltage mode:
If calibrating on the XFC: AIs support voltage mode by default. Ensure that there are no
external resistors connected across the AI pins.
If calibrating on the XRC: Verify that the mode selector jumper (J21, J22, J24, J25 or J26)
is across pins 2 and 3 for the selected AI.
To set the AI to current mode:
If calibrating on the XFC: Insert a precision resistor across the selected AI pins.
If calibrating on the XRC: Place the mode selector jumper (J21, J22, J24, J25 or J26)
across pins 1 and 2 for the selected AI.
11.5.3 Analog input calibration (voltage mode)
This procedure describes a three (3) point calibration. Available target values for this option display in
the calibration window and become active in calibration order:
Low Cal Point
100% Cal Point
50% Cal Point
These instructions can be adapted for a five (5) point calibration. The five pressure points display in
calibration order:
Low Cal Point
100% Cal Point
75% Cal Point
50% Cal Point
25% Cal Point

Table of Contents

Other manuals for ABB XFC G5

Related product manuals