54
Symbols, problems and error messages
12
Symbol Meaning
• The code chip has not been correctly inserted into the meter. Remove the code
chip and re‑insert it.
• There is no code chip in the meter. Insert the code chip that belongs to the
Accu‑Chek Active test strips you are currently using.
Averages cannot be calculated because
• you have not set the time and date,
• only test results without a time and date are saved,
• only test results that are not included in the calculation are saved, e.g. results
from control tests,
• no test results were saved in the period in question, e.g. all test results in
memory are more than 7 days old,
• the chronological order of the test results in the memory has been interrupted.
No test results have been saved.
mmol/L meter
mg/dL meter
Symbols appearing in the numeric field

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH