Display Shows Possible Cause(s) Possible Solution(s)
E-60 Time/Date
Error
j The time and date settings
may be incorrect.
j You have changed the
batteries.
j Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, “Changing Meter
Settings.”
HI bG Warning j Your blood sugar may be
higher than the measuring
range of the system.
j If you are experiencing any of the common
symptoms of high blood sugar, contact your
healthcare professional immediately.
j Treat your high blood sugar as recommended by
your healthcare professional.
j Consider checking bG, ketones, and insulin.
Insulin Sensitivity j Insulin sensitivity is outside of
the acceptable meter range.
j Check your entries and contact your healthcare
professional to determine the appropriate settings.
Invalid Bolus
Advice Times
j The acting time value is less
than the offset time.
j The acting time value must be set equal to or
greater than the offset time. Reset the acting time
value or revise the offset time.