EasyManua.ls Logo

Accu-Chek Aviva Expert - Page 244

Accu-Chek Aviva Expert
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
238
Display Shows Possible Cause(s) Possible Solution(s)
Invalid Date j The date entered is invalid
(dates for reminders cannot
be set to occur in the past).
j Re-enter date.
Invalid Hyper
Values
j The hyper warning limit value
must be greater than all of
your target ranges in the time
block settings.
j Reset the hyper warning limit or revise the target
ranges in time blocks and re-enter the hyper
warning limit.
j Enter a hyper warning limit that is above the target
ranges of your time blocks. See Chapter 5,
“Changing Meter Settings.
Invalid Hypo Values j The hypo warning limit value
must be less than all of your
target ranges in the time
block settings.
j Reset the hypo warning limit or revise the target
ranges in time blocks and re-enter the hypo
warning limit.
j Enter a hypo warning limit that is below the target
ranges of your time blocks. See Chapter 5,
“Changing Meter Settings.
45702_std.indb 238 10/29/09 1:42:52 PM