23
8. The result appears on the
display, along with the control
bottle symbol and a flashing
“L.” Do not remove the test
strip yet. Press and release
once to mark the result as a
Level 1. If you tested the
Level 2 control, press and
release
a second time.
9. Press and release
to set the
control level in the meter.
10. The control result and “OK”
alternate on the display if the
result is in range. The range is
printed on the test strip
container label. The control
result and “Err” alternate on
the display if the result is out
of range. Remove and discard
the used test strip.

Table of Contents

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Battery Type1 x 3V lithium (CR2032)
Operating Humidity10% to 90% RH
Meter Weight59 g
Measuring Range10 mg/dL to 600 mg/dL
Automatic Shut OffAfter 90 seconds of inactivity