Cleaning the meter and finger pricker
Wipe the measuring window and the area surrounding it with a cot-
ton swab lightly moistened with cold water or 70 % ethanol.
Remove any fluff or lint that may remain.
Allow areas you have wiped time to dry thoroughly.
Fold the drum compartment cover down and press it shut. There will
be an audible CLICK.
Perform a performance check (see Chapter 7).