Page #190 background imageLoading...
Page #190 background image
190
Symbols, error messages and troubleshooting
Error message Meaning
The batteries (regular or rechargeable)
are empty or a meter error has
occurred.
Turn the meter off and then on again
by pressing the Test button. If the error
message appears again, the batteries
may be empty. Insert new batteries
(new or recharged). If the error mes-
sage is again repeated when the
meter turns on, please contact the
Accu-Chek Customer Careline free on
0800 701000 (UK) or 1800 709600
(Ireland).
The meter was exposed to a strong
electromagnetic field.
Press the Test button to turn the meter
off. Move somewhere else or turn off
the source of the field.
You bent or moved the test strip during
the test either while you were drawing
up blood or control solution with it or
during the test.
쑺쑺
AC_CP_LCM_Manual_EN-UK final.indd 190_CP_LCM_Manual_EN-UK final.indd 190 20.6.2007 11:04:4620.6.2007 11:04:46

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH