Testing blood glucose in more than one patient
11
11.2 Disinfecting the meter
The following parts of the meter may become contaminated:
O the case
O the area around the test strip guide
The meter and the area around the test strip guide must be thoroughly
cleaned and disinfected at least once a day after use. Carefully check all
recesses, grooves and gaps.