Symbols, error messages and troubleshooting
13
This error message ... means:
A meter error has occurred.
Remove the test strip by pushing the
ejector down or turn the meter off. Start
again from the beginning. If the error
message appears again, the meter is
defective. Please contact your local cus-
tomer support and service centre (see
Chapter 19).
The ambient temperature or the tempera-
ture within the meter is too low or too
high for a test.
Remove the test strip by pushing the
ejector down or turn the meter off. Move
to a place where the temperature is
between +10 °C and +40 °C and wait for
the meter to warm up or cool down.