Testing blood glucose
Farbe: P 021 – Prüfmittelnummer: 201
Notes
Do not bend the test strip before it draws up blood or while it is
drawing up blood, and do not bend or move the test strip while
the test is in progress.
Bending or moving the test strip can produce an incorrect result
or cause an E-6 (see page 176) error message to be displayed.
Allowing the test strip to draw up blood too early, i.e. before the
flashing blood drop symbol has appeared, can produce an incor-
rect result or cause error message E-1 (see page 172) or E-6
(see page 176) to be displayed.
Remove the test strip by pushing the ejector down (see page 86)
and discard the test strip. Start the blood glucose test again with
a new test strip.
Always hold the meter so the test strip is pointing downwards
when you allow the test strip to draw up blood. If you hold it so
the test strip is pointing upwards or sideways or you lay the
meter down with the test strip still in, blood can enter the meter
and contaminate it.