41
Meter and Lancing Device Cleaning and Disinfecting
7
Chapter 7: Meter and Lancing Device Cleaning and Disinfecting
What is the dierence between cleaning and disinfecting?
Cleaning is the removal of dirt from the meter or lancing device.
3
Disinfecting is the removal of most, but not all, disease-causing and other types of microorganisms
(bloodborne pathogens) from the meter or lancing device.
3
Approved Cleaning and Disinfecting Product
The following product has been approved for cleaning and disinfecting the meter and lancing device:
Super Sani-Cloth (EPA*reg.no.9480-4)
Super Sani-Cloth can be purchased from Amazon.com, Ocedepot.com, and Walmart.com.
Do not use any other cleaning or disinfecting solutions. Using solutions other than the Super Sani-Cloth
could result in damage to the meter and lancing device.
The eect of using more than one product interchangeably to clean and disinfect the meter and lancing
device has not been tested. Always use Super Sani-Cloth to clean and disinfect the meter and lancing
device.
Roche has demonstrated that the product is good for 5-year use, after testing in a total of
260disinfection cycles (equal to cleaning and disinfecting once per week for 5years).
*Environmental Protection Agency
1080_Guide Me_MANUAL_MASTER.indb 41 3/5/2018 1:27:58 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Guide Me and is the answer not in the manual?

Accu-Chek Guide Me Specifications

General IconGeneral
DisplayLCD
Sample Size0.6 µL
Test TimeLess than 4 seconds
ConnectivityBluetooth
Measurement Range10 to 600 mg/dL
Battery Type2 x CR2032 3V lithium coin cells
Battery LifeApproximately 1000 tests
Weight40 g (1.4 oz) with battery
Storage Conditions-25°C to +70°C (-13°F to +158°F)