How to Clean and Disinfect the Meter
WARNING
Failure to follow these instructions will damage the meter and stop it
from working properly.
• DO NOT clean or disinfect the meter while performing a blood glucose
or control test.
•
DO NOT get any moisture in slots or openings.
• DO NOT spray anything onto the meter.
• DO NOT immerse the meter in liquid.
• Always use the same product for both cleaning and disinfecting.
Wash hands with soap and
water and dry thoroughly.
Turn the meter off and wipe the
entire meter surface with a
Super Sani-Cloth. Carefully
wipe around the test strip slot
and other openings.
Make sure that no liquid enters
any slot or opening.
A separate Super Sani-Cloth
should be used for cleaning and
disinfection. For disinfecting the
meter, get a new cloth and
repeat step2, making sure the
surface stays wet for 2minutes.
Make sure that no solution is
seen in any slot or opening.
1 2 3
1080_Guide Me_MANUAL_MASTER.indb 43 3/5/2018 1:27:59 PM