47
Meter Maintenance and Troubleshooting
8
Chapter 8: Meter Maintenance and Troubleshooting
Meter Maintenance
The meter automatically tests its own systems every time you turn it on and lets you know if something
is wrong. See the Error Messages section in this chapter.
If you have problems with the meter or think the results are not accurate, perform a control test with an
unexpired test strip and control solution. If the control result is not within the acceptable range, contact
the Accu-Chek Customer Care Service Center at 1-800-858-8072.
Changing the Batteries
Open the battery door on the
back of the meter by
pushing the tab in the
direction of the arrow and
pulling the door up.
1
SN XXXXXXXXXXX
PIN XXXXX
xxxxxxx, xxxxxxx
xxxx/x
GTIN XXXXXXXXXXXXXX
Remove the old batteries.
2
Slide the new batteries under
the black tabs with the (+)
side facing up.
Put the battery door back in
place and snap it closed.
3
NOTE
If the E-7 or E-9 error code still appears on the display after you have changed the batteries, remove
the batteries again, press for at least 2 seconds, then reinsert the batteries.
The meter uses two 3-volt lithium batteries, coin cell type CR2032. This type of battery can be found
in many stores. It is a good idea to have spare batteries available.
Always replace both batteries at the same time and with the same brand.
Meter data is not lost when you replace the batteries.
1080_Guide Me_MANUAL_MASTER.indb 47 3/5/2018 1:28:01 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Guide Me and is the answer not in the manual?

Accu-Chek Guide Me Specifications

General IconGeneral
DisplayLCD
Sample Size0.6 µL
Test TimeLess than 4 seconds
ConnectivityBluetooth
Measurement Range10 to 600 mg/dL
Battery Type2 x CR2032 3V lithium coin cells
Battery LifeApproximately 1000 tests
Weight40 g (1.4 oz) with battery
Storage Conditions-25°C to +70°C (-13°F to +158°F)

Related product manuals