2WIN / 2WIN-S
OPERATOR MANUAL

Table of Contents

Questions and Answers

  • R
    ralphhinesAug 19, 2025
    What to do if printing does not work on Adaptica 2WIN-S?
    • E
      Erica DavisAug 19, 2025
      If the Adaptica Measuring Instruments printer isn't working, first ensure the printer is powered on and has sufficient battery. If it is, make sure the printer is positioned within the line of sight of the instrument.
  • J
    Jessica MillerAug 21, 2025
    What to do if no symbols appear around the pupil on Adaptica 2WIN-S in OD/OS measurement mode?
    • R
      robertgilesAug 21, 2025
      If, in Normal Mode, no symbols appear around the pupil on your Adaptica Measuring Instruments, it could be due to the eyes not being within the measurement box, selecting the incorrect measurement mode (OD/OS), or both eyes being open. Ensure the eye is correctly positioned, the right mode is selected, and the patient covers the eye not being measured.
  • J
    Jacob WoodsAug 23, 2025
    How to fix a red circle around the pupil on Adaptica 2WIN-S Measuring Instruments?
    • K
      Keith DavisAug 23, 2025
      If you consistently see a red circle around the pupil(s) on the Adaptica Measuring Instruments, it means the patient isn't looking directly at the machine. If the patient is collaborative, ask them to focus on the device. If not, enable the interactivity features.
  • B
    Barbara DavisAug 26, 2025
    What to do if measurements with Adaptica Measuring Instruments are not working correctly?
    • K
      katrina16Aug 26, 2025
      If the measurement with Adaptica Measuring Instruments never completes correctly or quits prematurely while the pupils are detected, it might indicate a hardware problem. Run the Self Test diagnostic through the MENU and contact service.
  • T
    Tina SimmonsAug 27, 2025
    Why is there a red circle around the pupil even when the patient is looking at the Adaptica Measuring Instruments?
    • J
      James MeyerAug 27, 2025
      If you consistently see a red circle around the pupil(s) on the Adaptica Measuring Instruments even when the patient is looking at the machine (in Normal Mode), it may indicate a strabismus problem. The machine should advise about this condition after about 15 seconds.
  • W
    William SheltonAug 29, 2025
    How to fix Adaptica 2WIN-S Measuring Instruments when the initial screen does not appear?
    • J
      Jonathan WalshAug 30, 2025
      If the initial screen does not appear on the display of your Adaptica Measuring Instruments, the battery is completely discharged. Charge the battery.
  • P
    patriciablackwellSep 1, 2025
    What to do if Adaptica Measuring Instruments displays an error message due to software malfunction?
    • M
      Mr. Samuel MoralesSep 1, 2025
      If an error message appears on the display of your Adaptica Measuring Instruments due to software malfunctioning, contact the service center, mentioning the serial number and the specific error message.
  • H
    Hunter RobertsSep 3, 2025
    What to do if printing does not work with Adaptica Measuring Instruments?
    • C
      Chase SimpsonSep 3, 2025
      If the Adaptica Measuring Instruments printer isn't working, first ensure the printer is powered on and has sufficient battery. If that's not the issue, make sure the printer is positioned within the line of sight of the instrument.
  • J
    James WellsSep 5, 2025
    Why does 'Hint check Gaze' appear on Adaptica Measuring Instruments display and how to solve it?
    • S
      Shane TorresSep 5, 2025
      If you see a 'Hint check Gaze' warning on your Adaptica Measuring Instruments display, it means the patient isn't looking at the machine. If they are collaborative, ask them to focus on the device. If not, enable the interactivity features.
  • J
    Joseph HudsonSep 7, 2025
    What to do if a warning 'Possible strong Anisocoria' appears on the Adaptica Measuring Instruments display?
    • V
      Victoria WileySep 7, 2025
      If you see a 'Possible strong Anisocoria' warning on the Adaptica Measuring Instruments display, it indicates that the patient's eyes have very different pupil diameters. In this case, measure each eye individually in separate measurements.

Adaptica 2WIN-S Specifications

General IconGeneral
CategoryMeasuring Instruments
Refraction Measurement Range (Sphere)-15D to +15D
Data InterfaceUSB, Bluetooth
PowerRechargeable Battery
Resolution0.01D
Power SupplyRechargeable Li-ion Battery
ConnectivityUSB, Bluetooth
Astigmatism Axis1° to 180°
Fixation TargetInternal
Accuracy0.25D
Dimensions220mm x 320mm x 420mm
Operating Temperature10°C to 40°C
Storage Temperature-20°C to +50°C
Humidity30% to 80% (non-condensing)