It automatically measures binocular refraction, pupil distance, pupil size, direction of
gaze, and fixation abnormalities.
7.2 Main Applications
Refraction of infants and children from 2 months of age, seniors, impaired and non-
cooperative patients.
Early detection and documentation of multiple amblyogenic factors.
Un-aided binocular refraction of all other patients
Over-refraction of glasses or contact lenses.
OPTIONAL (APPs)
Phoria/tropia measurements
Dynamic pupil response time and size caused by artificial light created by
2WIN
Centering of the visual axis on the glasses frame
Addition power measurement at the 66 cm distance from the patient
7.3 Exam conditions
With reference to electromagnetic compatibility, the device is designed for use in a
Professional healthcare facility environment.
The patient shall sit comfortably in a suitable exam room; infants should sit in adult’s
lap. A uniform dim light environment is necessary, except in presence of the
KALEIDOS accessory.
The 2WIN accuracy is affected by external sources of light in particular HOT lights
halogen or incandescent lamps because of the presence of high infrared content of
the light; please make sure no direct light hits the patient’s face and ensure uniform
light on both eyes.
A stable measurement distance of 1 meter is also important.
7.4 Exam procedure
Turn the 2WIN ON by pressing the ON/OFF button for 1 sec. and wait until the stand-
by screen appears. A beep sound stands for 2WIN turning ON.
Please take care to modify the patient age before saving the 2WIN measurement.
The 2WIN results in young patient with hyperopia error can be dependant to the