46
Wait until the measurement process is finished and the results of the measurement
are displayed on the screen.
A two-beep sound indicates that the measurement has been correctly taken (this is
an option selectable from Measurement>Sound>Alerts>Ending Beep).
If the ambient light is too high a message will appear warning about the unsuitable
condition.
If something goes wrong during measurement, the measurement ends showing a
message indicating the cause of failure. These can be:
- Pupil/s too small or large;
- High IR noise;
- Unstable conditions of the data due to other eye problems;
- Other unexpected facts;
If this would happen try again. If the problem persists, suggest the patient to have a
specialist ocular examination.
If an error occurs during the measurement and the message appears, this last
message can be viewed by Notification Center.
If pupils diameter change during measurement, a message will be shown to display
after the measurement, please try to repeat the exam in stable light conditions.
Pupil diameter unstable
If in binocular refraction the two eyes have very different pupil diameters (that means
anisocoria) a popup will appear on the display. It is suggested to measure eyes
singularly in two different measurements.