28
Burst mode
It can be set to enable/disable, depending on whether the transmitter supports it.
Burst command
Burst command value can be set
Alarm status
Read-only parameter
(2) Operation
In the HART settings interface, you can view the HART parameter values in real time
After inputting the setting status of the current option, click to save
click to cancel saving and
return;
If the input value is displayed in red when saving, it means the input value is not accepted, please check its
input range;
If the current setting is canceled or the setting fails, the current item is restored.
5. Service
(1) Current loop test
Click to select diagnosis/service, enter the current loop test, and perform the current loop test. The value
range of this parameter is (4-20) mA;
The HART measurement value at the top of the calibrator interface is the actual value of the current loop.
(2) PV zeroing
Select zeroing in the HART diagnosis/maintenance interface, and enter the zeroing interface;
Ensure that the current measurement value is close enough to the zero point, otherwise it may cause zeroing
failure.
(3) Current adjustment
By adjusting the proportional coefficient of the current output link of the transmitter, the AO value of the

Related product manuals